Handyscope HS6 DIFF the USB oscilloscope packed with technology The new 1 GS/s 14 bit quad differential channel High Resolution Oscilloscope.

This powerful USB 3.0 super speed oscilloscope combines fast sampling up to 1 GS/s with high resolutions of 12, 14 and 16 bit and a large memory of 64 MSamples on all four channels. The oscilloscope supports continuous streaming measurements up to 200 MS/s and can be synchronized with other oscilloscopes using the CMI interface to form a multi channel combined instrument with synchronized time base. The CMI interface is available by default on the Handyscope HS6 DIFF. Optionally, the Handyscope HS6 DIFF can be delivered with SureConnect connection test and resistance measurement on each channel. Also, the Handyscope HS6 DIFF can be delivered with SafeGround option. With SafeGround you can switch the differential inputs of the Handyscope HS6 DIFF into single ended inputs with ground protection. It allows to make measurements using standard attenuating probes and protects the scope when a short circuit to ground is created.

Handyscope HS6 DIFF key specifications

This best in class High Resolution USB oscilloscope gives the best specifications available in the industry, for a limited budget. The flexibility and quality that the Handyscope HS6 DIFF offers is unparalleled by any other oscilloscope in its class.

Key specifications:

  • Channels: 4 Ch
  • Differential
  • SafeGround
  • Sample frequency: 1 GS/s
  • Record length: 256 MS
  • Resolution: 12 bit
  • Resolution: 14 bit
  • Resolution: 16 bit
  • Accuracy: 0.25%
  • Bandwidth: 250 MHz
  • Combine Multiple Instruments
  • USB3
  • RoHS
  • CD with software included
Oscilloscope / Spectrum analyzer / Multimeter / Data logger
14 bit (0.006 %) resolution (16 bit enhanced resolution)
Up to 1 GS/s sampling
250 MHz bandwidth
Up to 200 MS/s continuous streaming
Up to 256 MSamples memory per channel
0.25 % DC vertical accuracy, 0.1 % typical
1 ppm time base accuracy

The Handyscope HS6 DIFF can be synchronized with other oscilloscopes using the CMI interface to form a multi channel combined instrument with synchronized time base. The CMI interface is available by default on the Handyscope HS6 DIFF.

Optionally available: SafeGround: switch your differential inputs to single ended inputs, to use attenuating oscilloscope probes, with real time ground current protection on all channels.

Optionally available: SureConnect connection test: real time probe connection verification while measuring and fast and accurate resistance measurement on all channels.

Models

The Handyscope HS6 DIFF is available in 5 different base models:

Order code Max. sampling speed Max. streaming speed
HS6-DIFF-1000 1 GS/s 200 MS/s
HS6-DIFF-500 500 MS/s 100 MS/s
HS6-DIFF-200 200 MS/s 40 MS/s
HS6-DIFF-100 100 MS/s 20 MS/s
HS6-DIFF-50 50 MS/s 10 MS/s

Available options for the Handyscope HS6 DIFF are:

  • XM: With the extended memory option, up to 256 MSamples memory per channel is available. Without this option, the models have up to 1 MSamples memory per channel.
  • S: With the SureConnect option, connection test and resistance measurement are available on all channels.
  • G: with the SafeGround option, the differential inputs can be switched to single ended inputs, with a protection against shorts circuits to ground.
  • W5: With the extended warranty option, warranty is five years on parts and labor. Without this option, the models have three years warranty on parts and labor.

Package contents

The Handyscope HS6 DIFF models come with the following items:

Amount Item
1 Carry case BB391
1 Handyscope HS6 DIFF
4 Measure lead TP-C812B
4 Differential attenuator TP-DA10
1 Alligator Clip TP-AC80I - Set (8 clips: 4 black, 1 green, 1 blue, 1 yellow, 1 red)
1 Alligator Clip TP-AC10I - Set (8 clips: 4 black, 1 green, 1 blue, 1 yellow, 1 red)
1 Alligator Clip TP-AC5 - Set (8 clips: 4 black, 1 green, 1 blue, 1 yellow, 1 red)
1 Handyscope power supply
1 Probe compensation cable TP-DB9-BNC-30 (only in combination with option G SafeGround)
1 External power cable (USB)
1 CD-ROM with Multi Channel oscilloscope software and drivers
1 Instrument manual
1 Software manual

Handyscope HS6 DIFF features

 
  • SafeGround
  • SureConnect
  • Combining multiple Handyscope HS6 DIFFs
  • High accuracy
  • Large memory
  • Bandwidth limit
  • Many instruments

Differential / single ended switchable inputs with SafeGround protection

SafeGround SafeGround SafeGround gives the possibility to use the oscilloscope inputs both as single ended and as differential. When SafeGround is active and you accidentally create a wrong connection that causes a short circuit, SafeGround will disconnect the ground of the input channel without damaging the oscilloscope or PC. You can therefore simply switch from a differential input to a single ended input without worrying if anything will damage because of a short circuit current1. The Handyscope HS6 DIFF is the only oscilloscope in the world with this unique SafeGround protection. And as you all know, a connection mistake is easily made, which will now have no more strange and financial consequences because of SafeGround as the short circuit current is limited thanks to SafeGround.

Background: The advantage of an oscilloscope with differential inputs is that there are no connections between the channels and with the ground of the computer. It is therefore not possible to create a short circuit. With SafeGround enabled you can connect a standard 1:10 probe to your channel, this is not possible with a standard differential channel of other oscilloscope manufacturers. Sometimes it is also required to perform a single ended measurement, but then there is a risk of a short circuit.

When you do want to measure with a single ended input, your input is connected to the ground of your oscilloscope and your computer (the alligator clip of your probe is directly connected to ground). The input channels are also connected to each other. When you connect the alligator clip of your probe accidentally to a point in your test subject that is not at ground level but to a point with an elevated voltage, a short circuit current will flow through your probe, oscilloscope and computer. This can cause serious damage to the test subject, the scope and the computer. SafeGround avoids this and safes you a lot of misery. SafeGround can be enabled individually for each channel of the Handyscope HS6 DIFF.

SafeGround current limit menu The ground current at which the SafeGround protection activates and shuts off the ground connection is adjustable in steps of 10 mA. This allows to use single ended inputs in a situation where the instrument and the device under test do not have the exact same ground level, e.g. due to voltage drop over a ground connection. The small current that then will flow could activate SafeGround. Setting the current limit to a higher value may keep SafeGround from being activated.

When your device under test is very sensitive to damage, the SafeGround current limit can be set to a lower value, to give the highest level of protection.

SafeGround protects your scope, your computer and your circuit under test against accidental wrong ground connections.

SafeGround properties:

  • Low switch off current
  • High speed switching
  • High voltage protection
  • SafeGround on each channel

 

  1. Maximum short circuit current is 500 mA.

SureConnect connection test on each channel

SureConnect example While measuring, the revolutionary SureConnect connection test feature of the Handyscope HS6 DIFF checks in real time whether your test probe or clip actually makes electrical contact with your test subject.

Assuring a good connection of a probe with a test subject may not always be easy. The subject under measurement may be dirty, oxidized or an (invisible) protective layer may be present. Or, the test subject may be hidden, making visible contact confirmation impossible. Also, capacitive coupling between test probe and test subject can result in measuring a distorted version of the actual signal, wrongly suggesting a connection. Simply activate the SureConnect connection test and you know whether there is contact or not.

SureConnect: no more doubt whether your probe doesn’t make contact or there really is no signal.

As can be seen in the following video, capacitive coupling between test probe and test subject can result in measuring a distorted version of the actual signal, wrongly suggesting a connection. SureConnect however immediately shows you whether a good connection is present.

SureConnect is optionally available on the Handyscope HS6 DIFF.

Combining multiple Handyscope HS6 DIFFs for fully synchronized measuring

Combine Multiple Instruments The Handyscope HS6 DIFF is equipped with a sophisticated CMI synchronization bus, allowing to connect multiple Handyscope HS6 DIFFs to each other by means of coupling cables, to use them as a combined instrument. All instruments will measure at the same sample frequency (0 ppm deviation!). Apart from a synchronization bus, the CMI also contains a trigger bus and a detection bus. The maximum number of instruments is only limited by the number available USB ports.

Combining the instruments is fully automatic. When the Multi Channel oscilloscope software is started, the Handyscope HS6 DIFFs coupled with TP-C50H Coupling cables are identified and automatically combined to a larger instrument. Both the synchronization bus and trigger bus are automatically terminated at both ends with the correct impedance, so there is no need to manually place terminators. This unique possibility to create e.g. an 20 channel instrument is only available with the Handyscope HS6 DIFF and Handyscope HS5 and no other USB oscilloscope.

The video below shows how to combine and synchronize multiple oscilloscopes using the unique TiePie engineering Combine Multiple Instruments interface, where all channels are sampled using the same clock, with 0 ppm clock error, and shown together in one graph.
 

High amplitude resolution, 256 times more than a standard oscilloscope

Accuracy 0.25% The Handyscope HS6 DIFF measures with high resolutions of 14 and 16 bit. A signal measured with the Handyscope HS6 DIFF therefore has 256 times more resolution than most standalone oscilloscopes, which usually have a low resolution of 8 or 9 bit. The high resolution of the Handyscope HS6 DIFF precision oscilloscope allows for measuring signals with more accuracy, because the quantization error is much lower.

The input circuit of the Handyscope HS6 DIFF has a high accuracy of 0.25 % of the full scale input range, where most standalone oscilloscopes have an accuracy of 2 - 3 % of the full scale range.

To display a signal measured with the Handyscope HS6 DIFF High Resolution oscilloscope at the same level of detail as the standalone oscilloscope, the display can be 256 times larger. Viewing the signals on a 24" monitor immediately gives a very detailed impression of the signal. The smallest deviations are very well visible and because of the high resolution, it is still possible to zoom in and reveal additional details.

Handyscope HS6 high visible accuracy compared to competition.Figure 1: Handyscope HS6 high visible accuracy compared to competition.

Shown in figure 1 are two displays, both showing a measurement of the same signal. The left display size corresponds to a size comparable to a standalone oscilloscope; at 8 bit resolution, zooming will not reveal more details. The right display corresponds to a maximized window on a standard PC screen; at 14 bit resolution, zooming will still reveal more details.

The high resolution and high accuracy of the Handyscope HS6 DIFF and the unlimited zoom of the Multi Channel oscilloscope software allow to zoom in to mV levels. Figure 2 shows a measurement of a 2 mV sine wave made in the 200 mV input range, which is then zoomed in to view the signal at an 8 mV scale.

Zoomed in to get an 8 mV input scale.Figure 2: Zoomed in to get an 8 mV input scale.

Highly accurate 1 ppm oscilloscope time base

With a time base accuracy of 1 ppm, frequency and timing can be measured very accurately. The time base of the Handyscope HS6 DIFF is 25 to 100 times more accurate than comparable instruments of the competition.

Coupling multiple instruments to a large combined instrument does not affect the time base accuracy. The timing deviation between the coupled instruments is 0 ppm.
 

Mega deep memory of up to 256 MSamples per channel

256 MS memory When measuring at high sample rates, a long record length/large memory is necessary to be able to record a complete signal in the acquisition buffer. Where most oscilloscopes have 2.5 kSamples or 100 kSamples memory, the Handyscope HS6 DIFF has up to 256 MSamples memory per channel, depending on the selected resolution and the number of active channels. When measuring at 14 bit resolution and all four channels, the available memory is 32 MSamples per channel. This gives the user 300 to 100000 times more memory. An advantage of a large memory is that once-only fast phenomena can be captured accurately or complete serial communication signal blocks can be measured all at once. For example complete serial communications, like CAN bus signals, can be measured all in one record to be reviewed and analyzed afterwards.

The large memory results in the advantage of a large dynamic range in the USB spectrum analyzer, which enables troubleshooting in the frequency domain.

The unlimited super zoom feature of the Handyscope HS6 DIFF allows to zoom in to individual samples, no matter what record length is used.

The same signal viewed simultaneously with several zoom factors up to 1,000,000Figure 1: The same signal viewed simultaneously with several zoom factors up to 1,000,000

Figure 1 shows the same signal three times with different zooming factors. The lowest graph shows just 256 ns of the total 256 ms, a zoom factor of 1 million and still provides enough detail for accurate signal analysis.

Record lengths larger than 32 MSamples are only available in the 64 bit version of the Multi Channel oscilloscope software.

Switchable hardware-based bandwidth limiter

Handyscope HS6 DIFF Bandwidth Limit Handyscope HS6 DIFF Bandwidth Limit It seems reasonable to assume that more bandwidth is better, but a wider bandwidth gives more noise. To reduce your noise you can switch a bandwidth limiter on. When a lot of noise appears on your signal and triggering becomes unstable, switching the bandwidth limit on will give a stable triggering. Enabling the bandwidth limiter also avoids under sampling.

The bandwidth limit can be enabled for each channel individually on the channel toolbar and has four settings. The frequencies of the bandwidth limiter depend on the Handyscope HS6 DIFF model:

HS6 DIFF-1000 HS6 DIFF-500 HS6 DIFF-200 HS6 DIFF-100 HS6 DIFF-50
Off (250 MHz) Off (250 MHz) Off (250 MHz) Off (100 MHz) Off (100 MHz)
150 MHz 150 MHz 150 MHz 75 MHz 75 MHz
100 MHz 100 MHz 100 MHz 50 MHz 50 MHz
50 MHz 50 MHz 50 MHz 25 MHz 25 MHz

 

Many instruments in one device

  • PC based oscilloscope Oscilloscope
  • PC based multimeter Multimeter
  • PC based spectrum analyzer Spectrum analyzer
  • PC based data logger Data logger

With the Multi Channel oscilloscope software, the world's most sophisticated measuring software, the Handyscope HS6 DIFF can be used as: High Resolution USB Oscilloscope, multimeter, spectrum analyzer, data logger/transient recorder and protocol analyzer. This powerful and compact measuring instrument can solve almost any measurement problem.
 

Handyscope HS6 DIFF specifications

The tables below show detailed specifications of the Handyscope HS6 DIFF. Jump to acquisition specifications, trigger specifications or general specifications.

Acquisition system
Number of input channels 4 analog
CH1, CH2, CH3, CH4 Isolated male BNC
Type Differential input
Resolution 8, 12, 14, 16 bit user selectable
DC Accuracy 0.25 % (0.1 % typical) of full scale ± 1 LSB at 20°C to 25°C
To achieve rated accuracy, allow the instrument to settle for 20 minutes. When subjected to extreme temperatures, allow additional time for internal temperatures to stabilize.
Ranges ±200 mV to ±80 V full scale
Coupling AC/DC
Impedance 2 MOhm / 12 pF ± 1 %
1 MOhm / 20 pF ± 1 % when SafeGround enabled
Maximum voltage 200 V (DC + AC peak < 10 kHz)
Maximum Common Mode voltage 200 mV to 800 mV ranges : 2 V
2 V to 8 V ranges : 20 V
20 V to 80 V ranges : 200 V
Common Mode Rejection Ratio -47 dB
Bandwidth HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
-3dB at 75 % of full scale input 250 MHz 250 MHz 250 MHz 100 MHz 100 MHz
Limit, selectable per channel Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (250 MHz)
150 MHz
100 MHz
50 MHz
Off (100 MHz)
75 MHz
50 MHz
25 MHz
Off (100 MHz)
75 MHz
50 MHz
25 MHz
AC coupling cut off frequency (-3dB) ±1.5 Hz
SureConnect Optionally available (option S)
Maximum voltage on connection 200 V (DC + AC peak < 10 kHz)
Resistance measurement Optionally available (option S)
Ranges 100 Ohm to 2 MOhm full scale in 1-2-5 sequence
Accuracy 1 % of full scale
Response time (to 95%) < 10 µs
SafeGround Optionally available (option G)
Maximum voltage on connection 200 V (DC + AC peak < 10 kHz)
Maximum switch off current 500 mA
Response time < 100 ns
   
Maximum sampling rate HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
8 bit  
measuring one channel 1 GS/s 500 MS/s 200 MS/s 100 MS/s 50 MS/s
measuring two channels 500 MS/s 200 MS/s 100 MS/s 50 MS/s 20 MS/s
measuring three or four channels 200 MS/s 100 MS/s 50 MS/s 20 MS/s 10 MS/s
12 bit  
measuring one channel 500 MS/s 200 MS/s 100 MS/s 50 MS/s 20 MS/s
measuring two channels 200 MS/s 100 MS/s 50 MS/s 20 MS/s 10 MS/s
measuring three or four channels 100 MS/s 50 MS/s 20 MS/s 10 MS/s 5 MS/s
14 bit 100 MS/s 50 MS/s 20 MS/s 10 MS/s 5 MS/s
16 bit 6.25 MS/s 3.125 MS/s 1.25 MS/s 625 kS/s 312.5 kS/s
   
Maximum streaming rate 1 HS6D-1000 HS6D-500 HS6D-200 HS6D-100 HS6D-50
8 bit  
measuring one channel 200 MS/s 2 100 MS/s 2 40 MS/s 20 MS/s 10 MS/s
measuring two channels 200 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring three or four channels 100 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
12 bit  
measuring one channel 200 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring two channels 100 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
measuring three or four channels 50 MS/s 5 12.5 MS/s 5 5 MS/s 2.5 MS/s 1.25 MS/s
14 bit  
measuring one channel 100 MS/s 3 50 MS/s 3 20 MS/s 10 MS/s 5 MS/s
measuring two channels 50 MS/s 4 25 MS/s 4 10 MS/s 5 MS/s 2.5 MS/s
measuring three or four channels 25 MS/s 5 12.5 MS/s 5 5 MS/s 2.5 MS/s 1.25 MS/s
16 bit 6.25 MS/s 6 3.125 MS/s 1.25 MS/s 625 kS/s 312.5 kS/s
   
Sampling clock source  
Internal TCXO
Accuracy ± 0.0001 %
Stability ± 1 ppm over 0°C to 55°C
Time base aging ±1 ppm/year
External LVDS, on CMI connectors
Input range 10 MHz ± 1 %
16.369 MHz ± 1 %
   
Memory Standard model XM option
8 bit resolution  
Measuring one channel 1 MS per channel 256 MS per channel
Measuring two channels 512 KS per channel 128 MS per channel
Measuring three or four channels 256 KS per channel 64 MS per channel
12, 14, 16 bit resolution  
Measuring one channel 512 KS per channel 128 MS per channel
Measuring two channels 256 KS per channel 64 MS per channel
Measuring three or four channels 128 KS per channel 32 MS per channel
   
Trigger  
System Digital, 2 levels
Source CH1, CH2, CH3, CH4, OR, digital external
Trigger modes Rising edge, falling edge, any edge, inside window, outside window, enter window, exit window, pulse width
Level adjustment 0 to 100 % of full scale
Hysteresis adjustment 0 to 100 % of full scale
Resolution 0.006 % (14, 16 bits) / 0.025% (12 bits)
Pre trigger 0 to selected record length, 1 sample resolution
Post trigger 0 to selected record length, 1 sample resolution
Trigger hold-off 0 to 63 MSamples, 1 sample resolution
Trigger delay 0 to 16 GSamples, 1 sample resolution
Digital external trigger  
Input Extension connector pins 1 and 2
Range 0 to 2.5 V (TTL)
Coupling DC
Jitter ≤ 1 sample
 
Multi-instrument synchronization
Maximum number of instruments Limited by number of available USB ports
Synchronization accuracy 0 ppm
CMI interface 2x, CMI 1, CMI 2
Required coupling cable TP-C50H Coupling cable
Maximum coupling cable length 50 cm
 
Probe calibration
Output Extension connector pins 3 (signal) and 6 (ground)
Signal Square wave
Level -1 V to 1 V
Frequency 1 kHz
 
Interface
Interface USB 3.0 SuperSpeed (5 Gbit/s);
(USB 2.0 HighSpeed compatible)
 
Power Requirements
Power From USB port or external input
Consumption 5 VDC 1200 mA max
External power From second USB port or power adapter
 
Physical
Instrument  
Height 25 mm (1 inch)
Length 170 mm (6.7 inch)
Width 140 mm (5.2 inch)
Weight 500 g (17.6 ounce)
USB cord length 1.8 m (70 inch)
 
I/O connectors
Channel 1, 2, 3, 4 Isolated BNC
USB Fixed cable with USB 3.0 type A connector, 1.8 m
Extension connector D-sub 9 pins female
Power 3.5 mm power socket
CMI I/O connectors 2 x HDMI type C socket
 
System requirements
PC I/O connection USB 2.0, USB 3.0 or USB 3.1
Operating system Windows 7/8/10, 32 and 64 bits
Linux (via LibTiePie SDK)
 
Environmental conditions
Operating  
Ambient temperature 20°C to 25°C (10°C to 40°C without specifications)
Relative humidity 10 % to 90 %, non condensing
Storage  
Ambient temperature -20°C to 70°C
Relative humidity 5 % to 95 %, non condensing
 
Certifications and Compliances
CE mark compliance Yes
RoHS Yes
EN 55011:2009/A1:2010 Yes
EN 55022:2006/A1:2007 Yes
EN 61000-6-1:2007 Yes
EN 61000-6-3:2007 Yes
 
Package contents
Instrument Handyscope HS6 DIFF
Measure leads 4 x Measure lead TP-C812B, Differential BNC -> 4 mm banana jack
Differential attenuator 4 x Differential attenuator TP-DA10
Alligator clips, large Alligator Clip TP-AC80I - Set, 8 large alligator clips
Alligator clips, medium Alligator Clip TP-AC10I - Set, 8 medium alligator clips
Alligator clips, small Alligator Clip TP-AC5 - Set, 8 small alligator clips
Accessories Handyscope power supply, external power cable for USB port, Probe compensation cable TP-DB9-BNC-30 (only with option SafeGround)
Software for Windows 7/8/10 on CD
Drivers for Windows 7/8/10 on CD
Manuals instrument manual and software user's manuals
color printed and digital on CD
 
Optional Accessories
Optional accessories for the Handyscope HS6 DIFF need to be ordered separately
Oscilloscope probe Oscilloscope Probe 1:1-1:10 - HP-3250I, only with SafeGround option G
Measure lead single ended Measure lead TP-BNCI-100, only with SafeGround option G
 
Customer service
TiePie engineering instruments are designed, manufactured and tested to provide high reliability. In the unlikely event you experience difficulties, the TiePie engineering instruments are fully warranted for three years.
This warranty includes All parts and labor (excluding probes)
No charge for return shipping
Long-term 7 year support
Upgrade to the latest software at no charge
Optionally available Extension to five year warranty, option W5
  1. On some computers, the highest streaming rates may not be available, due to computer restrictions.
  2. ≤ 40 MS/s when connected to USB 2.0
  3. ≤ 20 MS/s when connected to USB 2.0
  4. ≤ 10 MS/s when connected to USB 2.0
  5. ≤ 5 MS/s when connected to USB 2.0
  6. ≤ 3.125 MS/s when connected to USB 2.0

 

Handyscope HS6 DIFF accessories

Optional accessories for the Handyscope HS6 DIFF are:

Measure lead TP-BNCI-100

Measure lead TP-BNCI-100 The Measure lead TP-BNCI-100 is a 1 meter long shielded single ended BNC to 4 mm banana jacks measure lead. It is to be used with single ended oscilloscope inputs. The BNC connector at one end plugs directly on the instrument. The two other ends each feature a single 4 mm banana jack, on which application specific test points, clamps or probes can be plugged.

The Measure lead TP-BNCI-100 can only be used when the Handyscope HS6 DIFF is equipped with option G SafeGround and when SafeGround is enabled.

Oscilloscope Probe 1:1-1:10 - HP-3250I

Oscilloscope Probe 1:1-1:10 - HP-3250I The Oscilloscope Probe 1:1-1:10 - HP-3250I is a 1:1-1:10 selectable passive high impedance oscilloscope probe with 250 MHz bandwidth (1:10) designed and calibrated for use on oscilloscopes having an input impedance of 1 MOhm shunted by 20 pF. However, it may be compensated for use with an oscilloscope having an input capacitance of 10 to 30 pF.

The Oscilloscope Probe 1:1-1:10 - HP-3250I can only be used when the Handyscope HS6 DIFF is equipped with option G SafeGround and when SafeGround is enabled.

To perform a frequeny compensation of the Oscilloscope Probe 1:1-1:10 - HP-3250I on the Handyscope HS6 DIFF, the Probe compensation cable TP-DB9-BNC-30 is required. It is included with the Handyscope HS6 DIFF when option G SafeGround is ordered.

Other accessories

Several other accessories are available for use with the Handyscope HS6 DIFF. The most common related accessories, such as measure leads, oscilloscope probes and current clamps are listed here. Refer to the products page to find all TiePie products.

  • Differential probe SI-9002
    Differential probe SI-9002
  • Current clamp TP-CC80: 0 - 20 A and 0 - 80 A input ranges
    Current clamp TP-CC80
  • Current clamp TP-CC600: 0 - 200 A and 0 - 600 A input ranges
    Current clamp TP-CC600
  • Current Clamp TP-CC400: 0 - 40 A and 0 - 400 A input ranges
    Current Clamp TP-CC400
  • Measure lead TP-C812B: low noise differential BNC to banana measuring lead, 2 m
    Measure lead TP-C812B
  • Measure lead TP-C1812B: low noise differential BNC to banana measuring lead, 3 m
    Measure lead TP-C1812B
  • Back Probe TP-BP85: thin and flexible back probe
    Back Probe TP-BP85
  • Alligator Clip TP-AC80I
    Alligator Clip TP-AC80I
  • Test Probe TP-TP90 Set
    Test Probe TP-TP90 Set
  • Alligator Clip TP-AC50B Set
    Alligator Clip TP-AC50B Set
  • Accelerometer TP-ACC20
    Accelerometer TP-ACC20
  • Handyscope power supply
    Handyscope power supply
  • TP-C50H Coupling cable
    TP-C50H Coupling cable
  • Rubber protector TP-RP-HS: protector for Handyscope series USB oscilloscopes
    Rubber protector TP-RP-HS
  • Alligator Clip TP-AC10I
    Alligator Clip TP-AC10I
  • Alligator Clip TP-AC5
    Alligator Clip TP-AC5
  • Oscilloscope Probe 1:1-1:10 - HP-3250I
    Oscilloscope Probe 1:1-1:10 - HP-3250I
  • Measure lead TP-BNCI-100
    Measure lead TP-BNCI-100
  • Differential attenuator TP-DA10
    Differential attenuator TP-DA10
  • Carry case BB391
    Carry case BB391
  • Probe compensation cable TP-DB9-BNC-30
    Probe compensation cable TP-DB9-

 

 

 

The Handyscope HS6 DIFF is available with different combinations of maximum sampling speed, memory depth, SureConnect, SafeGround and extended 5 year warranty. All models are listed below and can be ordered directly from stock.
Order code Max. sampling speed Max. streaming speed Memory SureConnect SafeGround Warranty
EJHS6-DIFF-1000XMSG-W5 1 GS/s 200 MS/s 256 MS 5 years
EJHS6-DIFF-1000XMSG 1 GS/s 200 MS/s 256 MS 3 years
EJHS6-DIFF-1000XMS-W5 1 GS/s 200 MS/s 256 MS   5 years
EJHS6-DIFF-1000XMS 1 GS/s 200 MS/s 256 MS   3 years
EJHS6-DIFF-1000XMG-W5 1 GS/s 200 MS/s 256 MS   5 years
EJHS6-DIFF-1000XMG 1 GS/s 200 MS/s 256 MS   3 years
EJHS6-DIFF-1000XM-W5 1 GS/s 200 MS/s 256 MS     5 years
EJHS6-DIFF-1000XM 1 GS/s 200 MS/s 256 MS     3 years
EJHS6-DIFF-1000SG-W5 1 GS/s 200 MS/s 1 MS 5 years
EJHS6-DIFF-1000SG 1 GS/s 200 MS/s 1 MS 3 years
EJHS6-DIFF-1000S-W5 1 GS/s 200 MS/s 1 MS   5 years
EJHS6-DIFF-1000S 1 GS/s 200 MS/s 1 MS   3 years
EJHS6-DIFF-1000G-W5 1 GS/s 200 MS/s 1 MS   5 years
EJHS6-DIFF-1000G 1 GS/s 200 MS/s 1 MS   3 years
EJHS6-DIFF-1000-W5 1 GS/s 200 MS/s 1 MS     5 years
EJHS6-DIFF-1000 1 GS/s 200 MS/s 1 MS     3 years
EJHS6-DIFF-500XMSG-W5 500 MS/s 100 MS/s 256 MS 5 years
EJHS6-DIFF-500XMSG 500 MS/s 100 MS/s 256 MS 3 years
EJHS6-DIFF-500XMS-W5 500 MS/s 100 MS/s 256 MS   5 years
EJHS6-DIFF-500XMS 500 MS/s 100 MS/s 256 MS   3 years
EJHS6-DIFF-500XMG-W5 500 MS/s 100 MS/s 256 MS   5 years
EJHS6-DIFF-500XMG 500 MS/s 100 MS/s 256 MS   3 years
EJHS6-DIFF-500XM-W5 500 MS/s 100 MS/s 256 MS     5 years
EJHS6-DIFF-500XM 500 MS/s 100 MS/s 256 MS     3 years
EJHS6-DIFF-500SG-W5 500 MS/s 100 MS/s 1 MS 5 years
EJHS6-DIFF-500SG 500 MS/s 100 MS/s 1 MS 3 years
EJHS6-DIFF-500S-W5 500 MS/s 100 MS/s 1 MS   5 years
EJHS6-DIFF-500S 500 MS/s 100 MS/s 1 MS   3 years
EJHS6-DIFF-500G-W5 500 MS/s 100 MS/s 1 MS   5 years
EJHS6-DIFF-500G 500 MS/s 100 MS/s 1 MS   3 years
EJHS6-DIFF-500-W5 500 MS/s 100 MS/s 1 MS     5 years
EJHS6-DIFF-500 500 MS/s 100 MS/s 1 MS     3 years
EJHS6-DIFF-200XMSG-W5 200 MS/s 40 MS/s 256 MS 5 years
EJHS6-DIFF-200XMSG 200 MS/s 40 MS/s 256 MS 3 years
EJHS6-DIFF-200XMS-W5 200 MS/s 40 MS/s 256 MS   5 years
EJHS6-DIFF-200XMS 200 MS/s 40 MS/s 256 MS   3 years
EJHS6-DIFF-200XMG-W5 200 MS/s 40 MS/s 256 MS   5 years
EJHS6-DIFF-200XMG 200 MS/s 40 MS/s 256 MS   3 years
EJHS6-DIFF-200XM-W5 200 MS/s 40 MS/s 256 MS     5 years
EJHS6-DIFF-200XM 200 MS/s 40 MS/s 256 MS     3 years
EJHS6-DIFF-200SG-W5 200 MS/s 40 MS/s 1 MS 5 years
EJHS6-DIFF-200SG 200 MS/s 40 MS/s 1 MS 3 years
EJHS6-DIFF-200S-W5 200 MS/s 40 MS/s 1 MS   5 years
EJHS6-DIFF-200S 200 MS/s 40 MS/s 1 MS   3 years
EJHS6-DIFF-200G-W5 200 MS/s 40 MS/s 1 MS   5 years
EJHS6-DIFF-200G 200 MS/s 40 MS/s 1 MS   3 years
EJHS6-DIFF-200-W5 200 MS/s 40 MS/s 1 MS     5 years
EJHS6-DIFF-200 200 MS/s 40 MS/s 1 MS     3 years
EJHS6-DIFF-100XMSG-W5 100 MS/s 20 MS/s 256 MS 5 years
EJHS6-DIFF-100XMSG 100 MS/s 20 MS/s 256 MS 3 years
EJHS6-DIFF-100XMS-W5 100 MS/s 20 MS/s 256 MS   5 years
EJHS6-DIFF-100XMS 100 MS/s 20 MS/s 256 MS   3 years
EJHS6-DIFF-100XMG-W5 100 MS/s 20 MS/s 256 MS   5 years
EJHS6-DIFF-100XMG 100 MS/s 20 MS/s 256 MS   3 years
EJHS6-DIFF-100XM-W5 100 MS/s 20 MS/s 256 MS     5 years
EJHS6-DIFF-100XM 100 MS/s 20 MS/s 256 MS     3 years
EJHS6-DIFF-100SG-W5 100 MS/s 20 MS/s 1 MS 5 years
EJHS6-DIFF-100SG 100 MS/s 20 MS/s 1 MS 3 years
EJHS6-DIFF-100S-W5 100 MS/s 20 MS/s 1 MS   5 years
EJHS6-DIFF-100S 100 MS/s 20 MS/s 1 MS   3 years
EJHS6-DIFF-100G-W5 100 MS/s 20 MS/s 1 MS   5 years
EJHS6-DIFF-100G 100 MS/s 20 MS/s 1 MS   3 years
EJHS6-DIFF-100-W5 100 MS/s 20 MS/s 1 MS     5 years
EJHS6-DIFF-100 100 MS/s 20 MS/s 1 MS     3 years
EJHS6-DIFF-50XMSG-W5 50 MS/s 10 MS/s 256 MS 5 years
EJHS6-DIFF-50XMSG 50 MS/s 10 MS/s 256 MS 3 years
EJHS6-DIFF-50XMS-W5 50 MS/s 10 MS/s 256 MS   5 years
EJHS6-DIFF-50XMS 50 MS/s 10 MS/s 256 MS   3 years
EJHS6-DIFF-50XMG-W5 50 MS/s 10 MS/s 256 MS   5 years
EJHS6-DIFF-50XMG 50 MS/s 10 MS/s 256 MS   3 years
EJHS6-DIFF-50XM-W5 50 MS/s 10 MS/s 256 MS     5 years
EJHS6-DIFF-50XM 50 MS/s 10 MS/s 256 MS     3 years
EJHS6-DIFF-50SG-W5 50 MS/s 10 MS/s 1 MS 5 years
EJHS6-DIFF-50SG 50 MS/s 10 MS/s 1 MS 3 years
EJHS6-DIFF-50S-W5 50 MS/s 10 MS/s 1 MS   5 years
EJHS6-DIFF-50S 50 MS/s 10 MS/s 1 MS   3 years
EJHS6-DIFF-50G-W5 50 MS/s 10 MS/s 1 MS   5 years
EJHS6-DIFF-50G 50 MS/s 10 MS/s 1 MS   3 years
EJHS6-DIFF-50-W5 50 MS/s 10 MS/s 1 MS     5 years
EJHS6-DIFF-50 50 MS/s 10 MS/s 1 MS     3 years
 
ETAP Users - Listed by Industry Electrical Engineering Consulting Firms ETAP

工程咨询

ETAP Users - Listed by Industry Data Centers ETAP

数据中心

ETAP Users - Listed by Industry Oil & Gas ETAP

采油平台

ETAP Users - Listed by Industry Generation Plants ETAP

发电厂

ETAP Users - Listed by Industry Nuclear Generation ETAP

核电厂

ETAP Users - Listed by Industry Transmission & Distribution ETAP

输配电

ETAP Users - Listed by Industry Government & Military Facilities ETAP

政府及军事设施

ETAP Users - Listed by Industry Metals & Mining ETAP

冶金矿业

ETAP Users - Listed by Industry Manufacturing ETAP

制造业

ETAP Users - Listed by Industry Renewable Energy ETAP

可再生能源

ETAP Users - Listed by Industry Transportation ETAP

铁路运输系统

ETAP Users - Listed by Industry Education - Colleges & Universities ETAP

教育系统

Handyscope HS6 DIFF downloads

Several downloads for the Handyscope HS6 DIFF are listed below, including Multi Channel oscilloscope software, drivers and manuals.

Driver & Software

Description Version Remarks Size  
USB driver 8.1.5 Windows 7/8/10, 32 and 64 bit 3.63 MiB  
Multi Channel software 1.37.1 32 and 64 bit, Windows 7/8/10, including PDF manual 58.73 MiB  
Multi Channel DEMO 1.37.1 32 and 64 bit, Windows 7/8/10, including PDF manual 57.30 MiB  

Manuals & Documents

Description Version Remarks Size  
Multi Channel software manual 1.07 English 4.63 MiB  
Handyscope HS6 DIFF Instrument manual 2.18 English 1.54 MiB  
Datasheet for Handyscope HS6 DIFF   English 2.41 MiB  

Programming

Interfacing the Handyscope HS6 DIFF with third party software and developing own software requires the LibTiePie Software Development Kit. Detailed information on the LibTiePie SDK, downloads for Windows and Linux platforms and bindings and programming examples for various programming environments can be found in the LibTiePie pages.